Error codeE08
The handswitch of the HF instrument is pressed while switching on. The manual's first action is: “Release the handswitch of the HF instrument.”
- Model
- Olympus ESG-400
- Display shows
- Handswitch pressed (E08)
- Type
- Error code
- Source
- Manual p. 122
What E08 means
Handswitch pressed p. 122
Common causes
- The handswitch of the HF instrument is pressed while switching on. p. 122
- Malfunction of the HF instrument. p. 122
- Malfunction of the electrosurgical generator. p. 122
Safety warnings — read before you start
What to do
- Release the handswitch of the HF instrument. p. 122
- Replace the HF instrument. p. 122
- If the problem persists, contact Olympus. p. 122
If E08 is still not resolved
This page summarises what the manufacturer’s manual says about E08; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.