Error codeE019
The footswitch combination error E019 indicates two single or double pedal footswitches are connected or the generator is malfunctioning. The manual's first action is: “Ensure that only one single pedal and / or only one double pedal footswitch are connected.”
- Model
- Olympus ESG-400
- Display shows
- Footswitch combination (E019)
- Type
- Error code
- Source
- Manual p. 122
What E019 means
Footswitch combination p. 122
Common causes
- Two single pedal or two double pedal footswitches have been connected. p. 122
- Malfunction of the electrosurgical generator. p. 122
What to do
- Ensure that only one single pedal and / or only one double pedal footswitch are connected. p. 122
- If the problem persists, contact Olympus. p. 122
Before you service the product, read the safety instructions in the manual p. 9.
If E019 is still not resolved
- If the problem persists, contact Olympus. p. 122 “If the problem persists, contact Olympus.”
This page summarises what the manufacturer’s manual says about E019; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.