Olympus ESG-400 Error E019: Footswitch combination

Electrosurgical generator · taken from the manufacturer’s manual, pages cited

Error codeE019

The footswitch combination error E019 indicates two single or double pedal footswitches are connected or the generator is malfunctioning. The manual's first action is: “Ensure that only one single pedal and / or only one double pedal footswitch are connected.”

Model
Olympus ESG-400
Display shows
Footswitch combination (E019)
Type
Error code
Source
Manual p. 122

What E019 means

Footswitch combination p. 122

Common causes

  • Two single pedal or two double pedal footswitches have been connected. p. 122
  • Malfunction of the electrosurgical generator. p. 122

Safety warnings — read before you start

  • DANGERDANGER. Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. p. 9
  • WARNINGWARNING. High frequency leakage current or spark discharge may cause user and / or patient burns. p. 9
  • CAUTIONCAUTION. Injury during servicing When the housing is opened, there is a danger of electric shock. The electrosurgical generator must only be serviced by authorized technicians. p. 12

What to do

  1. Ensure that only one single pedal and / or only one double pedal footswitch are connected. p. 122
  2. If the problem persists, contact Olympus. p. 122

If E019 is still not resolved

  • If the problem persists, contact Olympus. p. 122
    “If the problem persists, contact Olympus.”

This page summarises what the manufacturer’s manual says about E019; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.