Error codeE002
The system detected a short circuit in the HF instrument or its connection. The manual's first action is: “Ensure that the electrodes of the HF instrument do not touch each other.”
- Model
- Olympus ESG-400
- Display shows
- Short circuit (E002)
- Type
- Error code
- Source
- Manual p. 121
What E002 means
Short circuit p. 121
Common causes
- Electrodes of the HF instrument may touch each other. p. 121
- Malfunction of the HF instrument and / or the connection cable. p. 121
- Malfunction of the electrosurgical generator. p. 121
What to do
- Ensure that the electrodes of the HF instrument do not touch each other. p. 121
- Replace the HF instrument and / or the connection cable. p. 121
- If the problem persists, contact Olympus. p. 121
Before you service the product, read the safety instructions in the manual p. 9.
If E002 is still not resolved
- If the problem persists, contact Olympus. p. 121 “If the problem persists, contact Olympus.”
This page summarises what the manufacturer’s manual says about E002; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.