Olympus ESG-400 Error E002: Short circuit

Electrosurgical generator · taken from the manufacturer’s manual, pages cited

Error codeE002

The system detected a short circuit in the HF instrument or its connection. The manual's first action is: “Ensure that the electrodes of the HF instrument do not touch each other.”

Model
Olympus ESG-400
Display shows
Short circuit (E002)
Type
Error code
Source
Manual p. 121

What E002 means

Short circuit p. 121

Common causes

  • Electrodes of the HF instrument may touch each other. p. 121
  • Malfunction of the HF instrument and / or the connection cable. p. 121
  • Malfunction of the electrosurgical generator. p. 121

What to do

  1. Ensure that the electrodes of the HF instrument do not touch each other. p. 121
  2. Replace the HF instrument and / or the connection cable. p. 121
  3. If the problem persists, contact Olympus. p. 121

Before you service the product, read the safety instructions in the manual p. 9.

If E002 is still not resolved

  • If the problem persists, contact Olympus. p. 121
    “If the problem persists, contact Olympus.”

This page summarises what the manufacturer’s manual says about E002; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.