Olympus ESG-400 Error E394: Communication error

Electrosurgical generator · taken from the manufacturer’s manual, pages cited

Error codeE394

Communication error indicates improper cable connection or cable damage. The manual's first action is: “Check the connection of the communication cables to the electrosurgical generator and / or to the peripheral equipment for correct connection.”

Model
Olympus ESG-400
Display shows
Communication error (E394)
Type
Error code
Source
Manual p. 124

What E394 means

Communication error p. 124

Common causes

  • Improper connection of the communication cables to the LINK-OUT / LINK-IN socket on the rear panel of the electrosurgical generator or to the peripheral equipment. p. 124
  • Malfunction or damage of the communication cables. p. 124
  • Malfunction of the electrosurgical generator. p. 124

Safety warnings — read before you start

  • DANGERDANGER. Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. p. 9
  • WARNINGWARNING. Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. p. 9
  • CAUTIONCAUTION. Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. p. 9

What to do

  1. Check the connection of the communication cables to the electrosurgical generator and / or to the peripheral equipment for correct connection. p. 124
  2. Check the cables for damages and, if necessary, replace the cables. p. 124
  3. Contact Olympus. p. 124

If E394 is still not resolved

  • Contact Olympus. p. 124
    “Contact Olympus.”

This page summarises what the manufacturer’s manual says about E394; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.