Error codeE001
The HF instrument or its connection may have no proper tissue contact or be malfunctioning. The manual's first action is: “Ensure that the electrodes of the HF instrument have proper tissue contact.”
- Model
- Olympus ESG-400
- Display shows
- Open circuit (E001)
- Type
- Error code
- Source
- Manual p. 121
What E001 means
Open circuit p. 121
Common causes
- Electrodes of the HF instrument may have no proper tissue contact. p. 121
- Malfunction of the HF instrument and / or the connection cable. p. 121
- Malfunction of the electrosurgical generator. p. 121
Safety warnings — read before you start
- DANGERDANGER. Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. p. 9
- WARNINGWARNING. High frequency leakage current or spark discharge may cause user and / or patient burns. p. 9
- CAUTIONCAUTION. Injury during servicing When the housing is opened, there is a danger of electric shock. The electrosurgical generator must only be serviced by authorized technicians. p. 12
What to do
- Ensure that the electrodes of the HF instrument have proper tissue contact. p. 121
- Replace the HF instrument and / or the connection cable. p. 121
- If the problem persists, contact Olympus. p. 121
If E001 is still not resolved
- If the problem persists, contact Olympus. p. 121 “If the problem persists, contact Olympus.”
This page summarises what the manufacturer’s manual says about E001; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.