Error codeE187
The high frequency leakage current has exceeded the limit for monopolar or bipolar application. The manual's first action is: “Check if the electrosurgical generator or the patient is grounded unintentionally.”
- Model
- Olympus ESG-400
- Display shows
- Increased HF leakage current (E187)
- Type
- Error code
- Source
- Manual p. 124
What E187 means
Increased HF leakage current p. 124
Common causes
- The high frequency leakage current has exceeded the limit of 150 mA for monopolar application or 100 mA for bipolar application. p. 124
- Malfunction of the electrosurgical generator. p. 124
Safety warnings — read before you start
- DANGERDANGER. High frequency leakage current or spark discharge may cause user and / or patient burns. p. 9
What to do
- Check if the electrosurgical generator or the patient is grounded unintentionally. p. 124
- If the problem persists, contact Olympus. p. 124
If E187 is still not resolved
- If the problem persists, contact Olympus. p. 124 “If the problem persists, contact Olympus.”
This page summarises what the manufacturer’s manual says about E187; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Olympus. Found a mistake? Report an error.