Fluke 5626/5628 Platinum Resistance Thermometer User's Guide covering safety, specifications, installation, PRT care, operation and troubleshooting.
This user's guide covers the Fluke 5626 and 5628 Platinum Resistance Thermometers (PRT), secondary standard instruments that convert temperature to resistance. The 5626 has a resistance of 100 Ω and the 5628 has 25.5 Ω; both cover the ITS-90 range from −200 °C to 661 °C and are used with a readout device to detect temperature changes or actual temperature. The document was published in March 2014.
The guide opens with symbols and safety information, then gives specifications, construction and electrical circuit details. It continues with installation (environmental issues, mounting, lead wire identification), PRT care and handling guidelines, operation (comparison calibration, immersion requirements, thermal EMF), accessories such as case options and PRT termination, and a troubleshooting chapter. The PRTs are calibrated in Hart's Scientific Calibration laboratory and supplied with a NIST traceable Report of Calibration.
| Temperature range | −200 °C to 661 °C |
|---|---|
| Nominal resistance, 5626 | 100 Ω ±1 Ω at 0.01 °C |
| Nominal resistance, 5628 | 25.5 Ω ±0.5 Ω at 0.01 °C |
| Calibration uncertainty (k=2) | ±0.006 °C at −200 °C; ±0.004 °C at 0 °C; ±0.009 °C at 420 °C; ±0.014 °C at 661 °C |
| Short-term stability | ±0.003 °C (5626), ±0.002 °C (5628) |
| Long-term drift (Rtp) | ≤ 0.03 °C (5626) / ≤ 0.02 °C (5628) with 500 hours at 661 °C |
| Sheath | Inconel 600, 6.35 mm (0.25 in) diameter |
| Handle temperature range | 0 °C to 80 °C |
| Lead wires | 4-wire Super-Flex PVC, 22 AWG |
| Termination | Gold-plated spade lugs |
| Standard lengths | 12, 15 and 20 inches (304.8 mm, 381 mm, 508 mm) |
The two models use different standard resistance values: the 5626 is 100 Ω and the 5628 is 25.5 Ω.
They cover the International Temperature Scale of 1990 (ITS-90) range from −200 °C to 661 °C.
Recalibration should be scheduled according to your company's Quality Assurance requirements; normally a PRT is recalibrated annually.
It is intended for calibration work only and must not be used for any other application.