About this manual

This manual provides detailed safety precautions, unpacking instructions, and installation guidelines for the Thermo Scientific Revco PLUS upright ultra‑low temperature freezers. It explains proper location, wiring, leveling, and the optional pressure‑equalization port and remote alarm contacts.

The guide also describes the control panel functions, start‑up procedures, temperature setting, alarm configuration, and routine maintenance such as cleaning the condenser, gasket care, and defrosting. Troubleshooting tips and chart recorder options are included for reliable long‑term operation.

What's inside

  • About Ultra‑Low Temperature Freezers
  • Safety Precautions
  • Unpacking
  • General Recommendations
  • Operating Standards
  • Installation
  • Control Panel
  • Start‑Up and Operation
  • Backup System (Optional)
  • Maintenance and Troubleshooting

Specifications

Operating EnvironmentIndoor use
Maximum Altitude2000 m
Maximum Relative Humidity60 % (up to 32 °C)
Supply Voltage Fluctuation±10 % of nominal voltage

Frequently asked questions

What safety precautions must be followed before installing the freezer?

Read all safety warnings, ensure the unit is properly grounded, avoid modifying components, and verify the installation site meets clearance and temperature requirements.

How do I open and close the upright freezer door?

To open, remove any padlock, rotate the key counter‑clockwise, pull the latch handle toward you, then pull the door. To close, pull the latch to the open position, push the door shut, rotate the latch to engage, and lock with the key.

What are the environmental limits for operating the freezer?

The freezer can be used indoors up to 2000 m altitude, with a maximum relative humidity of 60 % at temperatures up to 32 °C, and ambient temperatures between 59 °F and 90 °F.

How can I set the cold and warm alarm setpoints?

With the key switch in the Power On position, press and hold the Cold (or Warm) Alarm Setpoint pad, then use the increment/decrement pads to adjust the value, ensuring a 5 °C difference from the cabinet setpoint.