RAE Systems ToxiRAE II PGM-1100 Series Personal Toxic Gas Monitor User Manual

User manual for the RAE Systems ToxiRAE II PGM-1100 Series personal toxic gas monitor, covering operation, calibration, alarms, maintenance and specs.


RAE Systems ToxiRAE II PGM-1100 Series Personal Toxic Gas Monitor User Manual - cover page
Brand
RAE Systems
Category
detectors
Document type
User Manual
Language
English
Pages
24
File format
PDF
File size
2 MB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
9676831BC34AE1E0E7270106A4C3AD59

About this manual

This is the user manual (045-4003-001, Rev A, December 2012) for the RAE Systems ToxiRAE II PGM-1100 Series personal toxic gas monitor. The ToxiRAE II is a single-gas monitor that continuously displays toxic gas concentrations together with STEL, TWA and Peak values, and it raises high, low, TWA and STEL alarms. The manual opens with standard contents and calibration kit lists, a general description, a physical description and an explanation of the display features.

It then explains how to turn the monitor on and off, how to read STEL, TWA and PEAK, and how to enter Program Mode and perform zero and span calibration or change preset limits and span gas values. A maintenance section covers changing the battery, the filter and the sensor, and the back of the manual gives specifications, Table A with sensor configurations and Table B with alarm signals for H2S, plus RAE Systems contact and service information.

What's inside

  • Standard Contents
  • Calibration Kit
  • General Information
  • Physical Description
  • Display Features
  • Operating the ToxiRAE II
  • Program Mode
  • Calibration
  • Changing Preset Limits or Span Gas Values
  • Maintenance

Specifications

Size3.6" (9.3 cm) x 1.9" (4.9 cm) x 0.9" (2.2 cm)
Weight3.6 oz (102 gm) with clip
BatteryUser-replaceable 2/3 AA high-capacity Lithium battery
Operating Period2 years typical battery life for CO/H2S/O2 sensors or 730 minutes of alarm; typical 1 year or 360 minutes of alarm for other sensors
DisplayEasy-to-read LCD for continuous display of concentration in ppm and text messages
Audible Alarm90dB at 12" (10 cm)
IP RatingIP-65, protected against dust and low pressure jets of water from all directions
Temperature-40° to 131° F (-40° to 55° C) for O2, CO, and H2S
Humidity0-95% relative humidity (non-condensing)
CalibrationTwo-point field calibration; auto-zero at startup (firmware version 1.4 only)
Warranty2-year warranty for instrument and CO/H2S/O2 sensors; 1-year for all other sensors

Frequently asked questions

How does the ToxiRAE II monitor turn on and off?

Hold the button down and release when the display says "on" to start it; hold the button through the "5…4… 3… 2… 1… OFF" sequence to turn it off, and the display goes blank.

How long before the STEL reading appears?

The STEL is the average reading for the last 15 minutes and does not appear until 15 minutes have elapsed.

How is the TWA calculated?

The TWA is the accumulated gas concentration reading since the monitor was turned on, divided by 8 hours.

Are the sensors interchangeable?

No. Only RAE Systems sensors of the type specified for your ToxiRAE II monitor may be used, and using non-RAE Systems components voids the warranty.