This manual describes how to use the QRAE II gas detector.
The RAE Systems QRAE II is a programmable multi-gas monitor designed for continuous exposure monitoring in hazardous environments. It features a user-friendly menu-driven interface with audio and visual alarms to detect oxygen, hydrogen sulfide, carbon monoxide, and combustible gases. The device is available in both diffusion and pumped models.
| Weight (Diffusion) | 9 oz (250 g) |
|---|---|
| Operating Time (Li-ion) | Up to 10 hours |
| Data Storage | 64,000 readings |
| Alarm Level | >95 dB @ 30 cm |
The monitor should be calibrated if it does not pass a bump test, but no less than every 6 months.
RAE Systems recommends using RAE calibration gas cylinders with a 4-gas mix containing 10 ppm H2S, 50 ppm CO, 50% LEL Methane, and 18% Oxygen.