Product data sheet for Micro Motion LNGS06S and LNGM10S LNG Series Coriolis meters, covering accuracy, flow ratings, pressure and approvals.
This product sample (PS-002080 Rev C, August 2019) presents the Micro Motion LNG Series meters for liquefied natural gas distribution applications. It explains the Coriolis measurement principle and the meter's mass flow, temperature and density measurement functions, along with configuration lock, data logging and diagnostics.
The document lists performance specifications for the LNGS06S and LNGM10S, including ±0.5% batch accuracy, ±0.25% repeatability, a –196 to +60 °C process medium temperature range and a 50 barg maximum sensor and process fitting pressure. It also covers hazardous-area certifications, electrical connections for the 820 core processor and Modbus/RS-485 communications. This manual is written in Chinese.
| Repeatability (LNGS06S / LNGM10S) | ±0.25% of batch |
|---|---|
| Process medium temperature range | –196 to +60 °C |
| Ambient temperature range | –40 to +60 °C |
| Maximum sensor and process fitting pressure | 50 barg |
| Nominal and maximum flow, LNGS06S | 900 kg/h nominal; 1800 kg/h maximum; 15:1 range |
| Nominal and maximum flow, LNGM10S | 11400 kg/h nominal; 18000 kg/h maximum; 15:1 range |
| Zero stability | LNGS06S: 0.6 kg/h; LNGM10S: 6 kg/h |
| Modbus/RS-485 data rates | 4800, 9600, 19200 and 38400 baud |
The document covers the LNGS06S and LNGM10S models.
The process medium temperature range is –196 to +60 °C.
The maximum working pressure is 50 barg.