Installation manual for Micro Motion F-Series Coriolis flow and density sensors covering planning, mounting, wiring, grounding, and case purge procedures.
This installation manual for the Micro Motion F-Series Coriolis Flow and Density Sensors covers planning, mounting, wiring, and grounding. It includes an installation checklist, best practices, temperature limits, cable length tables, and instructions for attaching extended electronics.
The manual also provides supplementary information on purging the sensor case and pressure relief, along with safety and approval information and the return policy. It notes that full product specifications are in the product data sheet and troubleshooting information is in the configuration manual.
| Maximum 9-wire cable length to 9739 MVD transmitter | 1,000 ft (305 m) |
|---|---|
| Maximum 9-wire cable length to all other MVD transmitters | 60 ft (18 m) |
| Maximum 4-wire power cable length, 18 AWG | 1,000 ft (305 m) |
| Maximum RS-485 signal cable length, 22 AWG or larger | 1,000 ft (305 m) |
| Electronics ambient temperature limits | Must not be below -40 °F (-40.0 °C) or above 140 °F (60.0 °C) |
The maximum length is 1,000 ft (305 m).
Yes. The sensor will work in any orientation as long as the flow tubes remain full of process fluid.
The electronics cannot be operated below -40 °F (-40.0 °C) or above 140 °F (60.0 °C).
The sensor case must be repurged whenever a purge plug is removed.