Micro Motion F-Series Coriolis Flow and Density Sensor Installation Manual

Installation manual for Micro Motion F-Series Coriolis flow and density sensors covering planning, mounting, wiring, grounding, and case purge procedures.


Micro Motion F-Series Coriolis Flow and Density Sensor Installation Manual - cover page
Brand
Micro Motion
Category
Mass Flowmeter
Document type
Installation Guide
Language
English
Pages
26
File format
PDF
File size
2.5 MB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
94B4462364622376073143691485B37D

About this manual

This installation manual for the Micro Motion F-Series Coriolis Flow and Density Sensors covers planning, mounting, wiring, and grounding. It includes an installation checklist, best practices, temperature limits, cable length tables, and instructions for attaching extended electronics.

The manual also provides supplementary information on purging the sensor case and pressure relief, along with safety and approval information and the return policy. It notes that full product specifications are in the product data sheet and troubleshooting information is in the configuration manual.

What's inside

  • Planning
  • Installation checklist
  • Best practices
  • Temperature limits
  • Mounting
  • Attach extended electronics
  • Wiring
  • Grounding
  • Purge the sensor case
  • Pressure relief

Specifications

Maximum 9-wire cable length to 9739 MVD transmitter1,000 ft (305 m)
Maximum 9-wire cable length to all other MVD transmitters60 ft (18 m)
Maximum 4-wire power cable length, 18 AWG1,000 ft (305 m)
Maximum RS-485 signal cable length, 22 AWG or larger1,000 ft (305 m)
Electronics ambient temperature limitsMust not be below -40 °F (-40.0 °C) or above 140 °F (60.0 °C)

Frequently asked questions

What is the maximum length of a Micro Motion 9-wire cable to a 9739 MVD transmitter?

The maximum length is 1,000 ft (305 m).

Can the F-Series sensor be installed in any orientation?

Yes. The sensor will work in any orientation as long as the flow tubes remain full of process fluid.

What are the ambient temperature limits for the electronics?

The electronics cannot be operated below -40 °F (-40.0 °C) or above 140 °F (60.0 °C).

What must be done whenever a purge plug is removed from the sensor case?

The sensor case must be repurged whenever a purge plug is removed.

Download PDF (2.5 MB)