Operating manual for FUT EK rotary torque sensors covering application, mechanical design, electrical connections, and maintenance procedures.
This document serves as the operating manual for the FUT EK Rotary Torque Sensors TRD/TRH/TRS 600/605/705 Series. It details the sensor's key features, including strain gage measuring systems and integrated rotation angle measurement capabilities for specific models. The guide explains the mechanical design variations between square and hexagon connections as well as round shaft options.
| Supply Voltage | 11 to 26 VDC |
|---|---|
| Calibration Supply | 4.5 to 30 VDC |
| Output Voltage | ± 5 VDC |
| Pulse Stripes | 360 light-dark stripes |
Yearly calibration is recommended, though this depends on the criticality of the application.
Bearing failure can be caused by torsional or flexural vibration, high axial or radial loads, worn bearings, or a bent shaft.
The signal lead should not exceed a length of 100 ft.