This document provides an application guide for camera measurement using Tektronix WFM5000, including camera adjustment and handling of white balance distortion.
This Tektronix application guide explains how to measure and correct camera white shading and white balance using a waveform monitor and a uniform light source. It describes setup with the DSC Ambi lighting equipment, the WFM7120 or portable WFM5000 waveform monitor, and camera HD-SDI output.
The guide covers lens and iris setup, RGB parade and vector displays, the diamond display for checking channel balance, and saving the resulting settings as a lens file preset for later use with the same lens and filter configuration. This manual is written in Chinese.
| Recommended waveform monitors | WFM7120, WFM7020, or portable WFM5000 |
|---|---|
| Recommended light source | DSC Ambi lighting equipment diffuser panel |
| Recommended illuminant | Industrial standard 3200 K tungsten source |
| Camera gain setting for measurement | 0 dB |
| Zebra coverage setting | 95% |
| Camera iris range | f4 to f5.6 |
| White balance color temperature after adjustment | 3200 K |
| Typical Ambi illumination deviation | +/- 0.5% |
| Vector display gain setting | 20x maximum |
The guide recommends an HD waveform monitor such as the WFM7120 or WFM7020, and also allows the portable WFM5000.
Use a very uniform light source such as the DSC Ambi lighting equipment, preferably with an industrial standard 3200 K tungsten source.
Set the camera gain to 0 dB and confirm that the knee and other gamma controls are turned off.
When the R, G, and B values are equal, a gray test chart or uniform white field appears as a vertical line in the upper and lower diamond displays, indicating correct white balance.