3M Clean-Trace ATP Monitoring System Bronchoscope Sampling Guide

Procedure guide for sampling Olympus flexible bronchoscopes with 3M Clean-Trace ATP tests to verify manual cleaning.


3M Clean-Trace ATP Monitoring System Bronchoscope Sampling Guide - cover page
Brand
3M
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
180 KB
Published
05 October, 2026
Updated
05 October, 2026
MD5 checksum
61226CEACEDAE00463729EED803ED889

About this manual

This guide describes the recommended procedure for obtaining acceptable ATP samples from Olympus® flexible bronchoscopes after manual cleaning. It outlines the materials, test points, and steps needed to assess residual clinical soil using 3M™ Clean-Trace™ ATP Surface and Water tests.

The document details how to prepare the bronchoscope, collect samples from the exterior surface and suction/biopsy channel, and interpret relative light unit (RLU) results. It also lists required equipment such as the Clean-Trace Luminometer, ATP test kits, syringes, and PPE, and provides reference FDA safety communications.

What's inside

  • Materials Required
  • Recommended Procedure for Sampling Olympus Model Flexible Bronchoscopes
  • Sample Exterior Surface
  • Preparation of Bronchoscope for Sampling of Suction/Biopsy Channel
  • Sampling Bronchoscope Channel
  • References

Specifications

RLU acceptance threshold200
Sample swab length10 cm
Sterile water volume drawn40 cc

Frequently asked questions

What RLU value indicates acceptable cleaning?

An RLU value below 200 is considered acceptable and allows the process to continue.

How far should the distal end of the bronchoscope be inserted when sampling?

The distal end should not go below the 40 mL mark in the collection tube to avoid contamination.