This document describes the preinstallation requirements for the Thermo EASY-nLC series.
Thermo Scientific EASY-nLC Series Preinstallation Requirements Guide covers preinstallation requirements for EASY-nLC II and EASY-nLC 1000 instruments. It includes an installation request form and checklist covering laboratory conditions, bench space, power, solvents, waste collection, telephone, and network access.
The guide also lists Thermo Fisher Scientific chromatography and mass spectrometry office contacts, regulatory compliance statements, and safety notices for lifting, handling, proper use, and electromagnetic susceptibility.
| Instrument weight (EASY-nLC II) | 32 kg (70.4 lb) |
|---|---|
| Instrument weight (EASY-nLC 1000) | 35 kg (77.0 lb) |
| Relative humidity | 20 to 80% (no condensation) |
| Remote support server | IP address 195.41.108.93, port 22 |
| Intended use | Research use only; not for diagnostic procedures |
EASY-nLC II and EASY-nLC 1000 instruments.
Relative humidity must be between 20 and 80% with no condensation.
The firewall must allow outgoing TCP/IP traffic to the remote support server at IP address 195.41.108.93, port 22.