ABB TBX556 TB(X)5 Series pH/ORP Sensor Quick Start Guide

The TBX556 manual introduces the TB(X)5 series of pH/ORP sensors for process monitoring.


ABB TBX556 TB(X)5 Series pH/ORP Sensor Quick Start Guide - cover page
Brand
ABB
Category
sensor
Document type
Quick Start Guide
Language
English
Pages
4
File format
PDF
File size
793 KB
Published
11 November, 2023
Updated
11 November, 2023
MD5 checksum
5E6E6344D5E1BBCE2C6289A144838FEB

About this manual

This quick start guide from ABB covers the TB(X)5 Series of pH/Redox (ORP) sensors for process monitoring, including the TB(X)556, TB(X)551, TB(X)557, TB(X)564 and TB(X)587 sensor styles. It explains how to install the sensors in pipelines and vessels, gives typical installation examples for retractable, twist-lock and threaded sensors, and lists the terminal block connections for TB82, TB84 and AX4xx analyzers.

The guide also describes integral cable (0 to 9 m / 0 to 30 ft) and extension cable (0 to 30 m / 0 to 100 ft) options, including BNC, Molex and optional VarioPin connections, and gives the recommended calibration procedure. A cleaning table lists cleaning solutions for different types of fouling, and a storage note explains how to keep the electrode from drying out.

What's inside

  • Installation
  • Electrical Connections
  • Typical Sensor Installations
  • Terminal Block Connections
  • Integral Cable
  • Extension Cable
  • Calibration
  • Cleaning
  • Storage

Specifications

Insertion Depth12 to 25 mm (0.5 to 1 in.)
Integral Cable Length0 to 9 m (0 to 30 ft.)
Extension Cable LengthUp to 30 m (100 ft.)
Junction Box Part Number (TB(X)5 with extension cable)4TB5023-0088
Junction Box Part Number (TB5 with extension cable)4TB5023-0162

Frequently asked questions

How should the TB(X)5 sensor be installed in a pipeline?

Install it in full pipelines to prevent the sensor from drying out, with the electrode pointing downwards.

How long can the integral cable be?

Integral cable can be specified from 0 to 9 m (0 to 30 ft.); ABB recommends keeping the length to a minimum for ease of sensor replacement.

When is an extension cable used?

An extension cable is used when the pH sensor is more than 9 m (30 ft.) but less than 30 m (100 ft.) from the analyzer.

How is the sensor calibrated?

Initial calibration must be a two-point calibration with liquid buffer solutions; later calibrations should be single-point grab-sample calibrations.

Download PDF (793 KB)