This manual introduces the basic operation, installation and maintenance of the 43iQHL High Level Sulfur Dioxide Analyzer.
The Thermo Scientific 43iQHL High Level SO2 Analyzer utilizes pulsed fluorescence technology to measure the amount of sulfur dioxide in the air up to 1000 ppm in standard range and up to 10000 ppm in extended range.
The manual covers unpacking and inspection, mounting options, setup and startup, instrument display and menus, calibration, data logging, settings, maintenance, troubleshooting, servicing, system description, optional equipment, safety, warranty, and WEEE compliance.
| Range | 0−1000 ppm |
|---|---|
| Extended Ranges | 0−10000 ppm |
| Zero Noise | 1 ppm RMS (10 second averaging time) |
| Detection Limit | 2 ppm (10 second averaging time) |
| Zero Drift | <1 ppm (24 hour) |
| Span Drift | ±1% full-scale (24 hour) |
| Response Time | 80 sec (10 second averaging time) |
| Linearity | ±1% full-scale |
| Flow Rate | 0.7 lpm (standard) |
| Operating Temperature Range | 0–45 °C |
| Power Requirements | 100–240 VAC 50/60 Hz 275 watts |
| Physical Dimensions | 24 in (D) x 16.75 in (W) x 8.72 in (H) |
| Weight | 37.1 lbs |
| Analog I/O | 4 Isolated Voltage Inputs 0–10 V |
| Digital I/O | 16 Digital Inputs (TTL) |
| Serial Ports | 1 RS-232/485 port |
| Other Ports | 3 Full Speed USB ports (one in front, two in rear) |
| Communication Protocols | MODBUS, Streaming, Bayern Hessen |
| Approvals and Certifications | CE, TUV-SUD Safety, UKCA |
It utilizes pulsed fluorescence technology.
The standard range is 0−1000 ppm.
The extended range is 0−10000 ppm.
It supports MODBUS, Streaming, and Bayern Hessen.