Thermo Scientific 43iQHL High Level Sulfur Dioxide Analyzer Instruction Manual

This manual introduces the basic operation, installation and maintenance of the 43iQHL High Level Sulfur Dioxide Analyzer.


Thermo Scientific 43iQHL High Level Sulfur Dioxide Analyzer Instruction Manual - cover page
Brand
Thermo Scientific
Category
Analyzers
Document type
User Manual
Language
English
Pages
268
File format
PDF
File size
15.2 MB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
23C5D5EB6483A7B0B20D56C02B02249E

About this manual

The Thermo Scientific 43iQHL High Level SO2 Analyzer utilizes pulsed fluorescence technology to measure the amount of sulfur dioxide in the air up to 1000 ppm in standard range and up to 10000 ppm in extended range.

The manual covers unpacking and inspection, mounting options, setup and startup, instrument display and menus, calibration, data logging, settings, maintenance, troubleshooting, servicing, system description, optional equipment, safety, warranty, and WEEE compliance.

What's inside

  • Introduction
  • Installation and Setup
  • Operation
  • Calibration
  • Maintenance
  • Troubleshooting
  • Servicing
  • System Description
  • Optional Equipment
  • Safety, Warranty, and WEEE

Specifications

Range0−1000 ppm
Extended Ranges0−10000 ppm
Zero Noise1 ppm RMS (10 second averaging time)
Detection Limit2 ppm (10 second averaging time)
Zero Drift<1 ppm (24 hour)
Span Drift±1% full-scale (24 hour)
Response Time80 sec (10 second averaging time)
Linearity±1% full-scale
Flow Rate0.7 lpm (standard)
Operating Temperature Range0–45 °C
Power Requirements100–240 VAC 50/60 Hz 275 watts
Physical Dimensions24 in (D) x 16.75 in (W) x 8.72 in (H)
Weight37.1 lbs
Analog I/O4 Isolated Voltage Inputs 0–10 V
Digital I/O16 Digital Inputs (TTL)
Serial Ports1 RS-232/485 port
Other Ports3 Full Speed USB ports (one in front, two in rear)
Communication ProtocolsMODBUS, Streaming, Bayern Hessen
Approvals and CertificationsCE, TUV-SUD Safety, UKCA

Frequently asked questions

What technology does the 43iQHL use to measure sulfur dioxide?

It utilizes pulsed fluorescence technology.

What is the standard measurement range of the 43iQHL?

The standard range is 0−1000 ppm.

What is the extended measurement range of the 43iQHL?

The extended range is 0−10000 ppm.

Which communication protocols does the 43iQHL support?

It supports MODBUS, Streaming, and Bayern Hessen.