This User Manual describes in detail the operation steps, precautions and possible abnormal situations of this product. Please read it carefully before using.
This Contec user manual describes a 12-lead electrocardiograph that samples ECG signals simultaneously and prints ECG waveforms using a thermal printing system. It covers intended use, main technical specifications, main characteristics and software overview.
The manual also provides safety precautions, working principle and part functions, operation precautions, preparation before operation, operation instructions and parameter setting, troubleshooting, maintenance, packing list and accessories, plus EMC guidance.
| Lead | Standard 12 leads |
|---|---|
| Input impedance | ≥2.5 MΩ |
| CMRR | >105 dB |
| Recording paper | 80 mm(W)×20 m(L) high-speed thermal paper |
| Paper speed | 12.5 mm/s, 25 mm/s, 50 mm/s, error ±5% |
| Gain control | 5, 10, 20 mm/mV, accuracy ±2% |
| Sampling precision | 24-bit |
| Battery | 7.4 V, 3500 mAh rechargeable lithium battery |
| Dimensions | 315 mm(L)×215 mm(W)×77 mm(H) |
| Net weight | 1.6 kg |
| Service life | 5 years |
It is intended for medical institutions to record human ECG signals and collect and extract the ECG waveform, including use in hospitals, wards and ambulances.
It uses 80 mm(W)×20 m(L) high-speed thermal paper.
It uses a 7.4 V, 3500 mAh rechargeable lithium battery.
The stated service life is 5 years.