Olympus 1X71 Optical Microscope User Manual

This is a user manual of Olympus 1X71 optical microscope


Olympus 1X71 Optical Microscope User Manual - cover page
Brand
Olympus
Document type
User Manual
Language
English
Pages
17
File format
PDF
File size
1.6 MB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
D51707245C9D78E78AFC0ED913CAB22B

About this manual

This manual provides instructions for operating the Olympus 1X71 optical microscope, including setup for transmitted light and epifluorescence modes. It details lamp operation, Kohler illumination adjustment, objective correction collar use, and fluorescence filter cube settings. The guide also covers image capture using the Hamamatsu ORCA-R2 camera and HClmage software, along with consumables and maintenance warnings.

What's inside

  • Ownership
  • Access Rules
  • General Specifications
  • Microscopy Techniques Available
  • Objectives
  • Filter Cubes
  • Camera
  • Fluorescence Illumination
  • Pixels to Microns Calibration
  • Consumables List

Specifications

Microscopy Techniques AvailableBrightfield, Phase contrast, Epifluorescence
ObjectivesOlympus UP lanFLN 10x Ph1 NA 0.3 WD 10.0 mm, Olympus LUCP lanFLN 20x Ph1 NA 0.45 WD 6.6 -7.8 with correction collar, Olympus LUCP lanFLN 40x Ph2 NA 0.6 WD 3.0 -4.2 with correction collar, Zeiss objectives can also be used
CameraHamamatsu ORCA R 2 CCD Camera with HClmage Capture Software
Fluorescence IlluminationPrior Lumen 200 Metal Halide Light Source (2000 hours/bulb)
Pixels to Microns Calibration5x objective binning 1x1 1 pixel = 1.88 um, 10x objective binning 1x1 1 pixel = 1.03 um, 20x objective binning 1x1 1 pixel = 0.514 um, 40x objective binning 1x1 1 pixel = 0.256 um

Frequently asked questions

How do I adjust the Kohler illumination?

Rotate the turret to the 'BF' position, open both iris diaphragms fully, focus the specimen, close the field iris diaphragm, adjust the condenser height, center the diaphragm, and open it until its image reaches the field of view limits.

What is the correct procedure for turning on the Halogen lamp?

Set both light intensity control knobs to MIN, turn the main switch to 'I' (ON) on the TH4 module, press the transmitted light ON-OFF button on the microscope, and adjust brightness using the control knob.

How do I adjust the objective correction collar?

Match the scale reading to the vessel bottom thickness if known, or rotate the collar left and right while refocusing to find the position yielding the best image resolution and contrast.

What are the warnings for using the Prior Lumen200 lamp?

Do not shut down the unit within 30 minutes of powering up or changing the bulb, as this may damage the bulb. Ensure the ventilation vent is unobstructed to prevent overheating.