This is a user manual of Olympus 1X71 optical microscope
This manual provides instructions for operating the Olympus 1X71 optical microscope, including setup for transmitted light and epifluorescence modes. It details lamp operation, Kohler illumination adjustment, objective correction collar use, and fluorescence filter cube settings. The guide also covers image capture using the Hamamatsu ORCA-R2 camera and HClmage software, along with consumables and maintenance warnings.
| Microscopy Techniques Available | Brightfield, Phase contrast, Epifluorescence |
|---|---|
| Objectives | Olympus UP lanFLN 10x Ph1 NA 0.3 WD 10.0 mm, Olympus LUCP lanFLN 20x Ph1 NA 0.45 WD 6.6 -7.8 with correction collar, Olympus LUCP lanFLN 40x Ph2 NA 0.6 WD 3.0 -4.2 with correction collar, Zeiss objectives can also be used |
| Camera | Hamamatsu ORCA R 2 CCD Camera with HClmage Capture Software |
| Fluorescence Illumination | Prior Lumen 200 Metal Halide Light Source (2000 hours/bulb) |
| Pixels to Microns Calibration | 5x objective binning 1x1 1 pixel = 1.88 um, 10x objective binning 1x1 1 pixel = 1.03 um, 20x objective binning 1x1 1 pixel = 0.514 um, 40x objective binning 1x1 1 pixel = 0.256 um |
Rotate the turret to the 'BF' position, open both iris diaphragms fully, focus the specimen, close the field iris diaphragm, adjust the condenser height, center the diaphragm, and open it until its image reaches the field of view limits.
Set both light intensity control knobs to MIN, turn the main switch to 'I' (ON) on the TH4 module, press the transmitted light ON-OFF button on the microscope, and adjust brightness using the control knob.
Match the scale reading to the vessel bottom thickness if known, or rotate the collar left and right while refocusing to find the position yielding the best image resolution and contrast.
Do not shut down the unit within 30 minutes of powering up or changing the bulb, as this may damage the bulb. Ensure the ventilation vent is unobstructed to prevent overheating.