Extech EC500 pH/Conductivity/SDT/Salinity/Temperature Meter User Manual

The ExStik® EC500 is a pH/Conductivity/SDT/Salinity/Temperature meter that supports multiple measurements, including pH, Conductivity, Total Dissolved Solids (TDS), and Salinity. It uses dynamic constant cell technology, making it possible to take a wide range of conductivity, TDS, and salinity measurements with the same electrode. This manual introduces you to how to use this meter.


Extech EC500 pH/Conductivity/SDT/Salinity/Temperature Meter User Manual - cover page
Brand
EXTECH
Document type
User Manual
Language
Spanish
Pages
13
File format
PDF
File size
568 KB
Published
18 October, 2023
Updated
18 October, 2023
Model
EC500
MD5 checksum
314411A674C86EC3201DEBB430B09AB9

About this manual

This manual provides instructions for operating the ExStik® EC500 meter, including calibration procedures, measurement modes, and battery management. It covers pH, conductivity, salinity, and temperature measurements with step-by-step guidance for setup and maintenance. The device uses four CR2032 lithium batteries and features automatic shutdown after 10 minutes of inactivity to conserve power.

The manual includes details on electrode care, calibration with standard solutions, and how to adjust units and settings. It also explains how to store and recall measurements, change temperature units, and reset calibration data. Users are advised to recalibrate before first use or after storage and to maintain electrode hydration for accurate readings. This manual is written in Spanish.

What's inside

  • Introduction
  • Battery and Power Management
  • Electrode Preparation and Care
  • Calibration Procedures
  • Measurement Modes
  • Temperature Unit Settings
  • Data Storage and Recall
  • Resetting Calibration Data
  • Troubleshooting

Specifications

Battery TypeFour (4) CR2032 Lithium
Automatic ShutdownAfter 10 minutes of inactivity
Calibration Points1, 2, or 3 point pH calibration
Temperature Units°C or °F
Conductivity Scale Factor0.4 to 1.0

Frequently asked questions

What should I do if the electrode indicator shows 'RENEW'?

Clean and recalibrate the electrode. If the indicator persists, replace the electrode as it no longer meets specifications.

What type of batteries does the EC500 use?

The EC500 uses four CR2032 lithium ion batteries.

How long does the EC500 stay on after inactivity?

The EC500 turns off automatically after 10 minutes of inactivity.

What is the range of conversion factors for SDT measurements?

The conversion factor for SDT ranges from 0.4 to 1.0.