This is a Kodak X-Omat 2000 PROCESSOR operator manual
The Kodak X-Omat 2000 and 2000A Processors are designed to process medical x-ray sheet or roll film using Kodak RP X-Omat Developer and Replenisher and Kodak RP X-Omat Fixer and Replenisher. The self-threading roller transport system features a film-length detection system, a replenishment system, a developer solution filter, and a recirculation system for the developer and fixer solutions. It also includes an automatic standby feature.
| Standard Cycle Transport Speed | 76.2 cm (30 in.) per minute |
|---|---|
| Standard Cycle Drop Time | 157 5 seconds |
| Standard Cycle Developer Temperature | 33.3 0.3 °C (92.0 0.5 °F) |
| Rapid Cycle Transport Speed | 101.6 cm (40.1 in.) per minute |
| Rapid Cycle Drop Time | 118 4 seconds |
| Rapid Cycle Developer Temperature | 34.4 0.3 °C (94.0 0.5 °F) |
The processor is designed to process medical x-ray sheet or roll film.
The recommended developer temperature is 33.3 °C (92.0 °F).
Only 3M SCOTCH Brand Polyester Film Tape No. 850 is acceptable.