EMERSON CA-3000/CA-3200 user manual

This document introduces medical product innovations since 1928, primarily the Emerson CoughAssist™ Mechanical In-Exsufflator (MI-E). The product assists patients in clearing retained bronchopulmonary secretions by gradually applying positive pressure to the airway, then rapidly shifting to negative pressure. It has automatic and manual models, which can be used with a facemask, mouthpiece, or endotracheal/tracheostomy tube.


EMERSON CA-3000/CA-3200 user manual - cover page
Brand
Emerson
Category
Other
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
599 KB
Published
10 August, 2012
Updated
28 September, 2023
MD5 checksum
0571239073C9D6EE252933026A60BFD4

About this manual

The Emerson CoughAssist™ Mechanical In-Exsufflator (MI-E) is a medical device designed to assist patients in clearing retained bronchopulmonary secretions. It operates by gradually applying positive pressure to the airway and then rapidly shifting to negative pressure to simulate a cough. The device is available in automatic models (CA-3000, CA-3200) and manual models (CM-3000, CM-3200), suitable for both tracheal and non-invasively ventilated patients.

What's inside

  • Introduction
  • Warnings and Cautions
  • Specifications
  • Controls, Connectors and Visual Indicators
  • Operating Procedure
  • Cleaning and Disinfection
  • Preventive Maintenance and Troubleshooting Guide
  • Warranty and Service

Specifications

Maximum Inhalation Flow3.3 liters/second
Maximum Exhalation Flow10 liters/second
Input Voltage (CA-3000, CM-3000)100-120 VAC, 60Hz
Input Voltage (CA-3200, CM-3200)220-240 VAC, 50Hz
Dimensions292mm x 279mm x 419mm (HxWxD)
Weight (CA-3200, CM-3200)9.3kg (20.6 lbs)

Frequently asked questions

How long can the device be cycled continuously?

The device is designed for intermittent operation only and should not be cycled continuously for more than 5 minutes.

What disinfectant is recommended for the device?

With the exception of the patient circuit, the device can be disinfected using 70% isopropyl alcohol or equivalent.

Can the device be used in the presence of flammable anesthetics?

No, the device is not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.

Download PDF (599 KB)