EXTECH EasyView Dual K Thermometer Model EA10 user manual

Introduces the features and usage of the Extech Datalogging Module 42270 and 42260 data loggers, including temperature, humidity and other measurement functions


EXTECH EasyView Dual K Thermometer Model EA10 user manual - cover page
Brand
EXTECH
Category
Other Electronics
Document type
User Manual
Language
English
Pages
1
File format
PDF
File size
72 KB
Published
01 August, 2012
Updated
30 September, 2023
Model
EA10
MD5 checksum
33F904914002757898985493205D413C

About this manual

This manual provides instructions for the Extech Model 42270 Temperature and Humidity Datalogging Module and the Model 42260 Temperature-only Datalogging Module. These devices are designed to monitor environmental conditions in various settings such as greenhouses, warehouses, food transports, aircraft cabins, refrigerated trucks, containers, museums, and HVAC equipment. The Model 42270 records both relative humidity and temperature, while the Model 42260 records temperature only.

What's inside

  • Introduction
  • Description
  • Recording Status Indication
  • ALARM Status Indication
  • Low Battery Indication
  • Specifications
  • Battery Replacement
  • Disposal

Specifications

Temperature range-40 to 85°C (-40 to 185°F)
Temperature accuracy±0.6°C (1.2°F) from -20 to 50°C
Humidity range0.0 to 99.9% Relative Humidity
Datalogger Memory16,000 temperature readings
Sampling rateProgrammable from 1 to 7200 seconds
Power supplyOne 3.6V ½ 'AA' lithium battery
Battery Life1 year: 5 second sample rate in sleep mode
Dimensions76.2 x 57.2 x 22.3mm

Frequently asked questions

What is the temperature range of the datalogger?

The temperature range is -40 to 85°C (-40 to 185°F).

How do I know if the battery is low?

The LCD displays 'LO' when the lithium battery voltage nears the critical operating level.

Can I use the datalogger without the software?

No, the dataloggers cannot be used without the software from kit 42275 (42265) or the printer programmer 42276 (42266).