Agilent Ultra-fast liquid chromatography using the Agilent 1100 Series HPLC system 1.8-um ZORBAX SB C18 Rapid Resolution HT columns Manual

This manual describes the use of 1.8-µm ZORBAX SB C18 Rapid Resolution HT columns with the Agilent 1100 Series HPLC system for ultra-fast liquid chromatography.


Agilent Ultra-fast liquid chromatography using the Agilent 1100 Series HPLC system 1.8-um ZORBAX SB C18 Rapid Resolution HT columns Manual - cover page
Brand
Agilent
Category
Other Electronics
Document type
Product Leaflet
Language
English
Pages
12
File format
PDF
File size
556 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
4A28E2AF3909E9518205D6517B50DE82

About this manual

This document describes the recommended system configuration to achieve very fast analysis and cycle times using ZORBAX 1.8-um Rapid Resolution High Throughput (RRHT) columns on the Agilent 1100 Series liquid chromatograph. It details how to reduce analysis time to seconds without compromising resolution or chromatographic performance.

What's inside

  • Technical Overview
  • Introduction
  • Instrumentation
  • Recommended System
  • Chromatographic Column
  • Automatic Delay Volume Reduction
  • Overlapped Injections

Specifications

Analysis time29 seconds
Peak width0.5 seconds
Backpressure300 bar
Column particle size1.8 um
Delay volume300 pL
Injection speed< 30 seconds
Flow cell volume500-nL

Frequently asked questions

What is the minimum configuration for ultra-fast analysis?

The minimum configuration consists of the Agilent 1100 Series binary pump, well-plate autosampler, thermostatted column compartment, and ZORBAX 1.8-um RRHT columns.

What is the benefit of overlapped injections (OD)?

Overlapped injection allows the preparation of the next injection while the current injection is being completed, increasing sample throughput.

Download PDF (556 KB)