Agilent 5973N Performance electronics: Considerations for GC-MS methods in scan selected-ion monitoring modes Manual

This note describes aspects of these improvements important to the successful implementation of more rapid gas chromatography/mass spectrometry analysis, in both scan and selected-ion monitoring modes.


Agilent 5973N Performance electronics: Considerations for GC-MS methods in scan selected-ion monitoring modes Manual - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
1.2 MB
Published
05 July, 2012
Updated
01 October, 2023
Model
5973N
MD5 checksum
A0422EFEAD8F4382EEFFA85304BAE9CD

About this manual

The document explains that higher productivity in gas chromatography requires faster mass spectrometric acquisition to capture information from narrower peaks. The new Performance Electronics of the Agilent 5973 Inert MSD were developed to maintain signal intensity and spectral quality while providing scan speeds up to 10,000 amu per second.

The text details how to select appropriate scan speeds based on whether the analysis is qualitative or quantitative. For quantitative studies requiring ion-ratio accuracy, 8 to 10 scans over a peak are necessary. For qualitative surveys where spectral quality is paramount, 4 to 5 scans are sufficient. The document provides charts and examples illustrating the relationship between mass range, sample numbers (averages), and effective scan speed.

What's inside

  • Scan Considerations
  • Quantitative Studies
  • Qualitative Studies
  • Effective Scan Speed
  • Sample Numbers
  • Mass Range
  • Detailed Example

Frequently asked questions

How many scans are necessary to establish ion ratios to approximately 1%?

Ten scans are necessary to establish ion ratios to approximately 1%.

What is the maximum scan speed of the 5973 Inert MSD electronics?

The electronics provide scan speeds up to 10,000 amu per second.

What is the recommended minimum number of scans over a peak for qualitative studies?

No less than 4 scans over a peak should be provided for qualitative studies.