HP 241 C-Terminal 5.0 Protein Sequencer Methods Technical Note

This technical note describes the C-Terminal 5.0 protein sequencer methods.


HP 241 C-Terminal 5.0 Protein Sequencer Methods Technical Note - cover page
Brand
HP
Category
Other
Document type
User Manual
Language
English
Pages
7
File format
PDF
File size
96 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
FF80DA05169BFF7C3CE2F4E63E78DF34

About this manual

This document describes the C-terminal 5.0 sequencer methods which implement thiohydantoin degradation chemistry on the Hewlett-Packard column-based protein sequencer. The methods are suitable for C-terminal sequence analysis of most protein samples in the low nanomole to 50 pmole range applied onto a Hewlett-Packard Zitex strip.

What's inside

  • C-Terminal 5.0
  • C-Terminal 5.0 (Cycle 1)
  • C-Terminal Flask 5.0
  • TH Std 5.0
  • Reagents
  • Method Details
  • Column Configuration

Specifications

Sample Rangelow nanomole to 50 pmole
Column Typeempty RP/SAX biphasic column
Strip TypeHewlett-Packard Zitex strip
Total Time (C-Terminal 5.0)75.2 min
Total Time (C-Terminal 5.0 Cycle 1)88.7 min

Frequently asked questions

What chemistry do the C-terminal 5.0 methods use?

They implement the thiohydantoin degradation chemistry (coupling and cleavage).

Are N-terminal sequencing reagents used for these methods?

No, reagents R1, R2A, R4, and the PTH-Std are specific for N-terminal sequencing and are not utilized.