Controlling Agilent 1200 Series RRLC Systems with Waters Empower Software Technical Overview

The Waters Empower 2 chromatography data software with ICS 1.05 fully supports Agilent 1200 Series Rapid Resolution LC (RRLC) systems with the exception of some special functionality. An Empower client-server system provides for easy installation and configuration of the Agilent 1200 Series RRLC system, support of all important parameters of the Agilent 1200 Series SL modules, evaluation of 2D, 3D and spectral raw data acquired by Agilent 1200 Series UV detectors, support of maintenance and diagnostics through Agilent Lab Monitor & Diagnostic software, and fast access to the Agilent 1200 Series RRLC system through the instant pilot controller.


Controlling Agilent 1200 Series RRLC Systems with Waters Empower Software Technical Overview - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
8
File format
PDF
File size
641 KB
Published
05 July, 2012
Updated
29 September, 2023
MD5 checksum
7177A27D07AEA25F12BBA9F69E4D87FD

About this manual

The document details the compatibility between Waters Empower 2 chromatography data software (with ICS 1.05) and Agilent 1200 Series Rapid Resolution LC (RRLC) systems. It outlines which specific Agilent LC modules are supported for control, noting exceptions such as certain pumps and autosamplers that require alternative configurations or are not fully controlled by Empower.

What's inside

  • Abstract
  • Introduction
  • Supported Agilent LC Modules
  • Non-Supported Agilent LC Configurations
  • Agilent Lab Monitor & Diagnostic Software
  • Instant Pilot Controller
  • Installation of the System
  • System Requirements
  • Setting the Communication Protocol

Specifications

Software Version TestedEmpower build 1154 SP L, Empower build 1154 SP M, Empower 2, FR3, Empower 2 FR4, Empower 2 SP D
Firmware Revisions TestedA.06.02/03/04, B.01.02/03/04 and B.02.01
Maximum Operating Pressureup to 600 bar
Maximum Data Rate (DAD SL)up to 80 Hz

Frequently asked questions

Which Empower software versions support Agilent LC ICS v1.05?

The supported versions include Empower build 1154 SP L, Empower build 1154 SP M, Empower 2, FR3, Empower 2 FR4, and Empower 2 SP D.

Can the Agilent 1200 Series RRLC be controlled by Waters Empower?

Yes, Waters Empower 2 with ICS 1.05 fully supports the system, except for some special functionality.