Agilent J&W Ultra Inert GC Columns Technical Overview

Describes the role of gas chromatography (GC) column quality control test probes (QC test probes), including ensuring that the columns have been properly deactivated, contain the correct amount of stationary phase, and have the same relative retention as the last column purchased.


Agilent J&W Ultra Inert GC Columns Technical Overview - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
4
File format
PDF
File size
258 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
25787194C6128F58DFE680A76A4C8F47

About this manual

This document introduces the Agilent J&W Ultra Inert GC column family, specifically the DB-5ms and HP-5ms models. It explains the importance of using specific Quality Control (QC) test probes to ensure column reproducibility and performance when analyzing chromatographically active analytes.

What's inside

  • Introduction
  • The Test
  • Probe Selection Is Critical
  • Less Demanding Test Mix Components
  • Uber One Mix
  • Conclusions

Frequently asked questions

What specific compounds are included in the Uber One QC test mix for evaluating column inertness?

The Uber One mix includes ten probes: 1-propionic acid, 1-octene, n-octane, 4-picoline, n-nonane, trimethyl phosphate, 1,2-pentanediol, n-propylbenzene, 1-heptanol, and 3-octanone.

What are the chromatographic conditions used to run the Uber One test mix on an Ultra Inert column?

The test uses a 65 °C isothermal oven, a split injection of 0.02 µL with 1 ng each component, and FID detection at 325 °C.

What column dimensions are specified for the Agilent J&W DB-5ms Ultra Inert column used in the successful test?

The column has dimensions of 30 m x 0.25 mm x 0.25 µm.