Agilent High Matrix Sample Introduction Accessory Performance Characteristics Manual for 7500 Series ICP-MS

This document introduces the Performance Characteristics of the Agilent High Matrix Sample Introduction (HMI) Accessory for the 7500 Series ICP-MS Product Overview, which can improve the performance of ICP-MS in high-concentration samples and reduce dissolved solids (TDS) levels to achieve better and more stable performance.


Agilent High Matrix Sample Introduction Accessory Performance Characteristics Manual for 7500 Series ICP-MS - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
6
File format
PDF
File size
245 KB
Published
05 July, 2012
Updated
01 October, 2023
MD5 checksum
A8B31FB2C2A582A23E9EC419A1EBBB53

About this manual

The High Matrix Sample Introduction (HMI) Accessory is designed for the Agilent 7500 Series Inductively Coupled Plasma Mass Spectrometer (ICP-MS). It utilizes an aerosol dilution technique combined with a robust Octopole Reaction System (ORS) to handle samples with high total dissolved solids (TDS), typically up to 1% or higher, without the need for conventional sample dilution. This approach eliminates associated drawbacks such as increased detection limits and contamination risks while significantly improving plasma stability.

What's inside

  • Introduction
  • Product Overview
  • How HMI Works
  • Sample Results
  • The Benefit of Lower Oxide Levels
  • Simple to Use
  • Applications of HMI

Frequently asked questions

What is the maximum TDS level the HMI accessory can handle?

The HMI accessory enables the 7500 ORS to analyze samples with up to 1% total dissolved solids (TDS) or higher, depending on the matrix.

How does HMI improve plasma robustness compared to conventional ICP-MS?

HMI reduces the CeO*/Ce* ratio down to approximately 0.2%, which is significantly lower than the typical 2% to 3% found in other instruments.

Does using HMI require disconnecting gas lines?

No, once the HMI is fitted, normal and HMI mode conditions can be switched automatically within a single sequence without disconnecting any gas lines.