Agilent Seamless method transfer from the Agilent 1200 Series to the Agilent 1220 1260 Infinity LC systems Manual

The document describes the process of seamlessly transferring validated methods from the Agilent 1200 Series LC system to the Agilent 1220 and 1260 Infinity LC systems. The method used USP/NF 23 method and the system suitability test developed on an Agilent 1200 Series LC system was used to test whether the Agilent 1220 and Agilent 1260 Infinity LC systems met the acceptance criteria. The chromatographic conditions of the system suitability test transferred to the Agilent 1260 and Agilent 1220 Infinity LC systems without changes. The results obtained were consistent with the results typically obtained on the Agilent 1200 Series LC system.


Agilent Seamless method transfer from the Agilent 1200 Series to the Agilent 1220 1260 Infinity LC systems Manual - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
8
File format
PDF
File size
224 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
978083795FC98999B4D126AEC44E90C6

About this manual

This document provides a technical overview of transferring validated chromatographic methods from the Agilent 1200 Series LC system to the newer Agilent 1220 Infinity and Agilent 1260 Infinity LC systems. It demonstrates that results obtained on the new instruments comply with acceptance criteria established for the original system without requiring method revalidation.

What's inside

  • Introduction
  • Instrumentation
  • Chromatographic conditions
  • Preparation of samples
  • Results and Discussion
  • Conclusion

Frequently asked questions

What is the maximum allowed difference in retention times between the new systems and the reference system?

The retention time shift for all peaks was less than 5% when using the same column and mobile phase batch.

What is the maximum acceptable relative standard deviation for peak area precision in USP methods?

The precision of areas must be less than 2% relative standard deviation.

What is the maximum acceptable relative standard deviation for retention time precision in USP methods?

The precision of retention times must be less than 0.5% relative standard deviation.