Agilent PLgel 5 um MIXED-C Eluent Modification Technical Overview

This document introduces the technical overview of Agilent PLgel 5 µm MIXED-C columns, including the adsorption phenomena in organic GPC separations and how to minimize this impact by modifying the eluent polarity.


Agilent PLgel 5 um MIXED-C Eluent Modification Technical Overview - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
559 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
929036388FB8A32CBCC3114441249536

About this manual

This document provides a technical overview regarding the use of eluent modification with Agilent PLgel 5 um MIXED-C columns. It addresses adsorption phenomena observed in organic GPC separations, specifically for polymer/solvent systems like polylactic acid in chloroform.

The text explains that modifying the eluent (e.g., adding methanol to chloroform) increases solvent polarity and minimizes solute/column interaction. The document confirms that PLgel columns are fully compatible with this modification, including water as a modifier up to about 10% by volume.

What's inside

  • Introduction
  • Eluent Modification
  • Adsorption Phenomena
  • PLgel Column Compatibility

Frequently asked questions

What is the maximum molecular weight limit for linear calibration on this column?

The PLgel 5 um MIXED-C columns provide a linear calibration up to 2 million MW.

What is the maximum percentage of water that can be used as an eluent modifier?

PLgel columns are fully compatible with water as a modifier up to about 10% by volume.

What is the specified flow rate for the analysis of polylactic acid?

The recommended flow rate for this analysis is 1.0 mL/min.