This document introduces the simple calibration method of Agilent OligoPore Columns (DMF), using the integrated Agilent PL-GPC 220 GPC/SEC system, using Agilent polyethylene glycol (PEG) narrow standards at 80 °C
This document provides a technical overview of the calibration procedure for Agilent OligoPore columns when used with dimethyl formamide (DMF) containing 0.1% LiBr. The process utilizes Agilent polyethylene glycol (PEG) narrow standards at an elevated temperature of 80 °C within the integrated PL-GPC 220 GPC/SEC system to ensure high resolution in the oligomeric region.
The calibration procedure uses a temperature of 80 °C.
The recommended flow rate is 1.0 mL/min.
An injection volume of 100 pL is required for the samples.