Agilent Cary 630 FTIR Specifications Datasheet

Agilent Cary 630 FTIR is a class-leading spectrometer for industrial applications as well as academic teaching and research support. Its ultracompact size, superior performance, versatility and ease of use make it an ideal choice for users that need a robust spectrometer for everyday, routine use.


Agilent Cary 630 FTIR Specifications Datasheet - cover page
Brand
Agilent
Category
Other Electronics
Document type
Datasheet
Language
English
Pages
2
File format
PDF
File size
778 KB
Published
04 July, 2012
Updated
01 October, 2023
MD5 checksum
A01251499453D6F3D76605ACA85CCAAB

About this manual

The Agilent Cary 630 FTIR is a class-leading spectrometer designed for industrial applications, academic teaching, and research support. It features an ultracompact design with a sealed and desiccated enclosure, making it robust for everyday routine use. The instrument utilizes a unique Flexture-based interferometer and solid-state laser for reliable, accurate, and repeatable results.

What's inside

  • Innovative design
  • Sample interface
  • Intuitive use
  • Technical specifications
  • Warranty
  • Hardware support
  • Software support

Specifications

Interferometer25 mm, permanently aligned, Michelson, 45°, mechanical flexture
EnclosureSealed and desiccated
Spectrometer interfaceUSB-2
Size16 x 31 x 13 cm (6 x 12 x 5 inches) (W x D x H)
Weight3.8 kg (8 lbs)
Power110 — 240 V AC, 60/50 Hz
Spectral range (KBr)7000-350 cm-1
Spectral range (ZnSe)5100-600 cm-1
Spectral resolution<2 cm-1
Wavenumber accuracy0.05 cm-1
Wavenumber reproducibility0.005 cm-1
Interferometer warrantyFive (5) year
Laser warrantyFive (5) year
Source warrantyThree (3) year

Frequently asked questions

What sample interfaces are available for the Agilent Cary 630 FTIR?

The instrument supports DialPath, TumblIR, Diamond ATR, Diffuse Reflectance, and Transmission.

Is the Agilent Cary 630 FTIR compliant with 21 CFR Part 11?

Yes, the optional Agilent MicroLab PC software is 21 CFR Part 11 compliant.

What is the spectral range of the Agilent Cary 630 FTIR with KBr optics?

The spectral range with KBr optics is 7000-350 cm-1.

Download PDF (778 KB)

More Agilent Manuals


View all Agilent manuals →