This document introduces the features and characteristics of Agilent ZORBAX Rapid Resolution High Definition Eclipse polynuclear aromatic hydrocarbon (PAH) threaded columns.
The Agilent ZORBAX Rapid Resolution High Definition Eclipse polynuclear aromatic hydrocarbon (PAH) threaded columns are designed for the high-efficiency separation of polynuclear aromatic hydrocarbons at higher pressure operation up to 1200 bar. The column packing is a polymeric C18 bonding designed to provide the high level of selectivity needed for PAH separations, utilizing a specially prepared, improved, ultra-high purity silica support.
| Maximum operating pressure | 1200 bar |
|---|---|
| Column dimensions | 2.1 x 100 mm |
| Surface area | 180 m²/g |
| Pore size | 95 Å |
| pH range | 2 to 8 |
| Maximum operating temperature | 60 °C |
The maximum operating pressure is 1200 bar (17,000 psi).
These columns are compatible with a pH range of 2 to 8.
Long-term storage should be in a pure organic solvent, preferably 100% acetonitrile.