Agilent 7510-AS Automated NMR Sample Changer Data Sheet

This document is the data sheet of Agilent 7510-AS NMR autosampler. This autosampler can be used for Agilent and Oxford magnets, which improves productivity and simplifies sample handling in your NMR laboratory. Each removable carrier holds up to 12 samples and supports random access. Load one carrier while running another. It is everything you expect in an autosampler: fast, smooth, and reliable operation that integrates with and simplifies your workflow without increasing maintenance. It is pneumatically-actuated and uses an optical encoder system to verify position while providing feedback to the controller.


Agilent 7510-AS Automated NMR Sample Changer Data Sheet - cover page
Brand
Agilent
Document type
Datasheet
Language
English
Pages
2
File format
PDF
File size
640 KB
Published
04 July, 2012
Updated
30 September, 2023
Model
7510-AS
MD5 checksum
D40D798EBB08E31129E22B0AE51D8290

About this manual

The Agilent 7510-AS is an automated NMR sample changer designed to improve productivity and simplify sample handling in NMR laboratories. It features a removable carrier holding up to 12 samples with random access capabilities and integrates seamlessly with Vnmrd software for intuitive operation.

The device utilizes pneumatically actuated mechanisms with optical encoders for position verification, ensuring reliable and smooth performance. It is compatible with various Agilent and Oxford magnet systems and requires minimal routine maintenance as there are no user-serviceable parts.

What's inside

  • Key Benefits
  • Introduction
  • NMR System Compatibility
  • Computer Requirements
  • Operating Conditions
  • Maintenance
  • Safety Certification

Specifications

Compatible sample tube diameter1-10 mm
Maximum sample length9 inches (7 inches for J-Young)
Samples per carrier12
Storage temperature-25 °C to 55 °C
Relative humidity (operating/storage)20 to 80 % non-condensing
Maximum altitude3000 m
Pneumatics pressure448-689.5 kPa (65-100 psig)
Driver module weight2.2 kg (4.8 lb)
Carrier weight (unloaded)2.8 kg (6.1 lb)

Frequently asked questions

Does the 7510-AS require routine maintenance?

No, it requires little routine maintenance and is designed to operate without adjustment.

What computer systems are compatible with the autosampler?

It is compatible with Dell 755, T3400, T3500 or HP 2400 computers.

Is the device CE compliant?

Yes, the 7510-AS is CE compliant and tested to EN61326-1 and FCC standards.