Agilent HPLC-Chip Intelligent Sample Loading Application Note

This Application Note demonstrates how Intelligent Sample Loading automates and optimizes sample enrichment on the HPLC-Chip, and also shows how Intelligent Sample Loading increases the number of peptides detected in a protein digest test mixture.


Agilent HPLC-Chip Intelligent Sample Loading Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
292 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
9D09FF22EF538CB9B83B8CC78E9689B7

About this manual

The document describes an application using the Agilent HPLC-Chip/MS system for automated sample enrichment. It highlights that conventional nanocolumn LC-MS systems require time-consuming manual adjustment of valve event timetables when injection volumes change, which can degrade efficiency.

Intelligent Sample Loading (ISL) is presented as a solution that decouples the sample loading process from analysis. The system automatically calculates the required loading time based on actual flow rates and capillary delay volumes, eliminating the need for manual re-optimization. This feature reduces data file sizes and increases the number of peptides detected in protein digest mixtures.

What's inside

  • Advantages of Intelligent Sample Loading
  • HPLC-Chip/MS System Components
  • Chromatographic Conditions
  • Workflow Description
  • Intelligent Sample Loading Parameters
  • Calculation of Injection Flush Volume

Specifications

Enrichment column40 nL, 5 um ZORBAX 300SB-C18
Analytical column75 pm x 43 mm, 5 pm ZORBAX 300SB-C18
Injection flush volume (IFV)5.926 pL

Frequently asked questions

How does Intelligent Sample Loading improve the enrichment process?

It automates and optimizes sample loading by automatically calculating the required time to effectively load the sample, eliminating the need for manual re-adjustment when injection volumes change.

What is the benefit of Intelligent Sample Loading regarding data files?

It reduces data file size because MS data acquisition coincides with the switching from enrichment to analysis mode instead of starting at the beginning of the autosampler sample loading cycle.