Agilent 6210/6510 TOF/QTOF Mass Spectrometry Application Note

This application note describes how to use the Agilent 1200 Series Rapid Resolution LC system and Agilent ZORBAX Rapid Resolution High Throughput columns to separate compounds in complex samples, how to calculate the empirical formula of unknown compounds from high mass accuracy data acquired with the Agilent 6210 time-of-flight and teh Agilent 6510 Q-TOF mass spectrometer, how to use the Agilent MassHunter Profiling software to quickly and computer-assistedly analyze complex mass spectrometry data, and how to use the Agilent MassHunter Profiling software to analyze the differences in related samples in different applications.


Agilent 6210/6510 TOF/QTOF Mass Spectrometry Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
4.9 MB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
6173417E2C948295B1A0B6B191857EE9

About this manual

This document describes a method for the differential analysis of complex samples, such as natural product extracts or drug metabolism reactions, using Agilent 6210 time-of-flight and Agilent 6510 quadrupole time-of-flight mass spectrometers. The analysis utilizes the Agilent MassHunter Profiling software to identify unknown compounds and examine differences between sample groups based on accurate mass data.

The study presents two case examples: the identification of pseudoginsenoside F11 in American ginseng versus Asian ginseng, and the detection of hydroxylated buspirone metabolites in a drug metabolism reaction. The text details the experimental setup, including the Agilent 1200 Series LC system and specific MS operating conditions, as well as the data processing workflow using Molecular Feature Extractor and Profiling software.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Method
  • Results and discussion
  • Identification of differences in complex natural product extracts
  • Identification of metabolites of a pharmaceutical drug
  • Table 1
  • Table 2

Specifications

Mass accuracy rangesingle-digit ppm mass error
Column typeAgilent ZORBAX SB C18, 2.1 x 150 mm, 1.8 pm particle size
Flow rate0.5 mL/min
Column temperature50 °C
DAD wavelength220 nm +4 nm, reference 360 nm + 16nm
Dry gas flow10 L/min
Dry Temp300 °C
Nebulizer pressure50 psi
Fragmentor voltage200 V
Skimmer voltage60 V
Capillary voltage3500 V
Scan range200-1300

Frequently asked questions

What software is used for the differential analysis?

The Agilent MassHunter Profiling software is used for the differential analysis of TOF and QTOF data.

Which mass spectrometers are used in the study?

The study uses the Agilent 6210 time-of-flight MS and the Agilent 6510 QTOF mass spectrometer.

What is the mass accuracy for pseudoginsenoside F11 in American ginseng?

The average relative mass accuracy for pseudoginsenoside F11 in the American ginseng sample group is 1.53 ppm.

How many molecular features were identified in the ginseng extract study?

671 molecular features were identified and displayed in the mass against retention time plot.