Agilent DB-XLB GC Analysis of Polybrominated Flame Retardants Application Note

Polybrominated diphenyl ethers (PBDE) are used as flame retardants in diverse products such as textiles, circuit boards, and computer covers. Through the disposal of those products in landfills, PBDEs have found their way into the environment. Studies have shown that PBDEs have detrimental health effects. This application note describes the development of suitable gas chromatography/mass spectrometry, gas chromatography micro electron capture detection, and gas chromatography inductively coupled plasma mass spectrometry methods to analyze PBDEs. The Agilent DB-XLB is the column of choice for this demanding analysis. The detection limit with micro electron capture detector was 100 ppt for most congeners.


Agilent DB-XLB GC Analysis of Polybrominated Flame Retardants Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
539 KB
Published
02 July, 2012
Updated
29 September, 2023
MD5 checksum
654972918D378FAA87CBF8E483D21DD3

About this manual

This document details the analysis of polybrominated diphenyl ethers (PBDEs), which are flame retardants found in the environment. It addresses the challenges of detecting these compounds due to their high boiling points and low thermal stability.

The note evaluates the Agilent DB-XLB column, a proprietary low-polarity stationary phase, alongside the DB-35ms column. It demonstrates that the DB-XLB provides excellent baseline separation of critical PBDE congeners in approximately 20 minutes using gas chromatography/mass spectrometry (GC/MS). The document also covers detection using micro electron capture detection (1ECD) and inductively coupled plasma mass spectrometry (ICP-MS).

What's inside

  • Introduction
  • Results and Discussion
  • GC/MS Analysis
  • Column Selection
  • Detection Methods
  • Test Mixtures
  • Chromatograms

Specifications

Detection limit (1ECD)100 ppt for most congeners
Carrier gasHelium at 38 cm/s
Oven temperature program100 C for 1 min; 100 C to 340 C at 20 C/min, 340 C for 12 min

Frequently asked questions

Which column is recommended for analyzing PBDEs?

The Agilent DB-XLB is the column of choice for this demanding analysis.

What detection modes were evaluated?

The study evaluated mass selective detector (MSD), micro electron capture detector (1ECD), and inductively coupled plasma mass spectrometry (ICP-MS).