This application note describes a method for the qualitative and quantitative analysis of explosives and related compounds using Agilent's ZORBAX®SB-C18 and SB-CN columns. The method combines the excellent separation capabilities of the polar and nonpolar columns and has been proven to perform well in a variety of applications.
This document details an application for the qualitative and quantitative analysis of explosives and related compounds using Agilent ZORBAX SB-C18 and ZORBAX SB-CN HPLC columns. The method utilizes a specific mobile phase composition and gradient elution profile to separate various nitro compounds.
| Column Dimensions | 2.1 x 150 mm |
|---|---|
| Column Part Number | 883700.905, 883700.922 |
| Injection Volume | 10 pL |
| Flow Rate | 0.23 mL/min |
| Temperature | 18C |
| Detection Wavelengths | 210, 240, 360 nm |
The application uses ZORBAX SB-C18 and ZORBAX SB-CN columns.
The recommended injection volume is 10 pL.
The analysis should be performed at 18C.