Agilent Analysis of 370-550 C Petroleum Fraction using a VF-5ht UltiMetal Column

This document describes the method of analyzing the 370-550 °C petroleum fraction using the Agilent J&W FactorFour VF-5ht UltiMetal column. The column has high temperature and corrosion resistance, which can improve the signal-to-noise ratio and detection limits.


Agilent Analysis of 370-550 C Petroleum Fraction using a VF-5ht UltiMetal Column - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
147 KB
Published
12 June, 2012
Updated
29 September, 2023
Model
VF-5HT
MD5 checksum
713788A5805380F91908580BBC861E42

About this manual

The document details the analysis of heavy hydrocarbon fractions (C20-C40) found in petroleum waxes and lube stocks. It recommends the on-column injection mode for this specific type of analysis due to the demanding conditions required.

The note provides specific chromatographic conditions for the analysis, including the use of a VF-5ht UltiMetal column (30 m x 0.25 mm, df=0.1 µm) with a hydrogen carrier gas. The temperature program starts at 40 °C and ramps to 400 °C, with peak identification provided for hydrocarbons ranging from C20 to C60.

What's inside

  • Introduction
  • Conditions
  • Peak Identification
  • Column Specifications
  • Detection Limits

Specifications

Column Length30 m
Column Inner Diameter0.25 mm
Film Thickness0.1 µm
Carrier GasHydrogen
Injector Temperature325 °C
Detector Temperature340 °C
Temperature Program40 °C (0.5 min) to 400 °C
Ramp Rate5 °C/min

Frequently asked questions

What is the recommended injection mode for 370-550 °C petroleum fractions?

On-column injection is the preferred mode for analyses of this type.

Which detector is used for the analysis described in this note?

The analysis uses a Flame Ionization Detector (FID) set to 340 °C.