Agilent PL HFIPgel Polyamide Analysis Application Note

This application note illustrates the use of Agilent PL HFIPgel columns for the analysis of a polyamide sample.


Agilent PL HFIPgel Polyamide Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
108 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
D5D183EBA6D48D480CC3C60958133F4D

About this manual

This document details the analysis of polyamide and polyester materials using Gel Permeation Chromatography (GPC) with Hexafluoroisopropanol (HFIP) as a solvent. The text highlights that HFIP is a favored solvent for these polymers due to its low refractive index and ability to operate at near ambient temperatures, unlike alternative solvents like m-cresol.

What's inside

  • Analysis of Polyamide
  • GPC System Setup
  • Calibration Curves
  • Chromatogram Results
  • Molecular Weight Distribution
  • Operating Conditions
  • Column Specifications
  • Conclusions

Specifications

Columns2 x PL HFIPgel, 300 x 7.5 mm
EluentHFIP + 0.2 M NaTFAc
Flow rate1.0 mL/min
Injection volume200 pL
Temperature40°C
DetectorRI
SystemAgilent PL-GPC 50

Frequently asked questions

What solvent is used for polyamide analysis in this note?

Hexafluoroisopropanol (HFIP) is used as the solvent.

Why is HFIP preferred over m-cresol or o-chlorophenol?

HFIP is used at near ambient temperatures and has a low refractive index compared to polyesters and polyamides.

What additive is added to the HFIP eluent?

0.02 M NaTFAc is added to minimize aggregation of the polymer in solution.