Agilent 1290 Infinity LC Application Note: Influence of Delay Volume and System Dispersion on Resolution and Sensitivity

This Application Note demonstrates the influence of system delay volume and system dispersion on the resulting resolution and sensitivity achieved for LC/MS experi- ments. The advantages of using a low delay volume LC system and optimized connec- tion of the LC to the MS are shown. The study illustrates that the Agilent 1290 Infinity LC system is an ideal partner for LC/MS instruments.


Agilent 1290 Infinity LC Application Note: Influence of Delay Volume and System Dispersion on Resolution and Sensitivity - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
806 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
ADBBABD1A20F4AA9510EDC25F63607FE

About this manual

This document investigates the impact of system delay volume and tubing dispersion on the performance of Liquid Chromatography-Mass Spectrometry (LC/MS) systems. It highlights that lower delay volumes and optimized connections are critical for maintaining high resolution and sensitivity, particularly for early-eluting compounds.

The study utilizes an Agilent 1290 Infinity LC system to compare separation results under standard low-delay conditions versus extended delay configurations. Additionally, it analyzes the effects of varying capillary inner diameters connecting the LC column outlet to the mass spectrometer source, demonstrating how larger diameters significantly degrade chromatographic resolution.

What's inside

  • Abstract
  • Introduction
  • Experimental Equipment and Methods
  • Influence of LC System Delay Volume
  • Influence of Connection to LC-Associated Detector
  • Results and Discussion

Frequently asked questions

What is the flow rate used for the HPLC method described in this application note?

The flow rate was set to 0.5 mL/min.

At what temperature was the column maintained during the experiment?

The column temperature was maintained at 40°C.