Agilent 1260 Infinity Binary LC Application Note for Quantification of Olaquindox and Carbadox

This Application Note describes a method for the quantifi cation of two antibacterial drugs, olaquindox and carbadox in animal feedstuff using the Agilent 1260 Infinity Binary LC system.


Agilent 1260 Infinity Binary LC Application Note for Quantification of Olaquindox and Carbadox - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
965 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
4E398B0AE3229A027EDD2515C22FB5BE

About this manual

This document presents a simple and sensitive analytical method for the routine monitoring of banned substances, specifically the growth promoters olaquindox and carbadox, in animal feedstuff. The method utilizes the Agilent 1260 Infinity Binary LC system equipped with a Max-Light 60 mm high-sensitivity flow cell and UV detection at specific wavelengths (375 nm for olaquindox and 306 nm for carbadox). The system configuration includes the G1312B Binary Pump, G1379B Degasser, G1367E Autosampler, G1330B Thermostat, and G1316B Column Compartment, controlled by ChemStation B.04.03 software.

What's inside

  • Abstract
  • Introduction
  • System
  • Chromatographic conditions
  • Samples
  • Results and discussion
  • Analysis of spikes samples
  • Conclusions
  • References

Specifications

ColumnAgilent ZORBAX SB-Aq 3.0 x 100 mm, 3.5 ym
Flow rate0.75 mL/min
Detection wavelength for olaquindox375 nm
Lower limit of quantification50 ng/mL
SoftwareChemStation B.04.03

Frequently asked questions

What is the lower limit of quantification for olaquindox and carbadox in this method?

The lower limit of quantification achieved for both molecules was 50 ng/mL.

What is the linear range for the area response of the analytes?

The area response was found to be linear up to 5,000 ng/mL.