Agilent 7700x ICP-MS Ultratrace Determination of Iodine-129 Application Note

This application note describes the ultratrace determination of iodine-129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode.


Agilent 7700x ICP-MS Ultratrace Determination of Iodine-129 Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
492 KB
Published
11 June, 2012
Updated
30 September, 2023
Model
7700x
MD5 checksum
5077C8E5F3E3152BAD92389B178F8D4C

About this manual

The document details a method for measuring the radionuclide iodine-129 (I-129) using Inductively Coupled Plasma Mass Spectrometry (ICP-MS). The study utilizes an Agilent 7700x ICP-MS equipped with an Octopole Reaction System (ORS) cell operated in oxygen reaction mode. This setup effectively removes background signals caused by Xenon impurities and polyatomic ions, enabling high-sensitivity detection without sample preparation.

What's inside

  • Introduction
  • Instrumentation
  • Optimization of oxygen flow rate
  • Calibration standards
  • Analytical data

Frequently asked questions

What is the detection limit for iodine-129 using this method?

A detection limit of 1.1 ppt was obtained for iodine-129.

Is sample preparation required for aqueous samples?

No sample preparation is required for aqueous samples.

What is the total cell gas flow rate at the optimum oxygen flow rate?

The total cell gas flow rate was 5.01 mL/min.