This application note shows the analysis of Aceton, n-Hexane, Methyl iso-Butyl Ketone (MIBK), Methyl iso-Butyl Carbinol (MIBC), and Methyl n-Butyl Ketone (MNBK, 2-Hexanone) using the Agilent 490 Micro GC.
This document details an environmental analysis method for monitoring soil decontamination. It focuses on analyzing a gas sample containing contaminants like Acetone, n-Hexane, Methyl iso-Butyl Ketone (MIBK), Methyl iso-Butyl Carbinol (MIBC), and Methyl n-Butyl Ketone (MNBK) using the Agilent 490 Micro GC.
The analysis utilizes a Nitrogen/Oxygen matrix equipped with CP-Sil 5 CB and CP-Wax 52 CB columns. The document provides specific chromatographic parameters, including column temperatures of 60°C, carrier gas flow (Helium at 200 kPa), and injection settings, along with typical results for the specified compounds.
| Product Part Number | G3581A |
|---|---|
| Column Temperature | 60°C |
| Carrier Gas Pressure | 200 kPa |
| Injector Temperature | 60°C |
| Injection Time | 200 ms |
The analysis covers Acetone, n-Hexane, Methyl iso-Butyl Ketone (MIBK), Methyl iso-Butyl Carbinol (MIBC), and Methyl n-Butyl Ketone (MNBK).
The system is equipped with a CP-Sil 5 CB column channel and a CP-Wax 52 CB column channel.