Agilent 5975T LTM GC/MSD Pesticide Detection Application Note

Rapid analysis of pesticides in vegetables by Agilent 5975T LTM GC/MSD and TSP


Agilent 5975T LTM GC/MSD Pesticide Detection Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
1.5 MB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
9708BC36FF5AA6688BFBF657B4A02282

About this manual

The document details an evaluation of the Agilent 5975T LTM GC/MSD for detecting pesticides in vegetable extracts without requiring a traditional clean-up step. It utilizes the Thermal Separation Probe (TSP) technology to minimize sample preparation time while maintaining quantitative and confirmatory results.

The study establishes a fast method using DRS software and RTL functions, achieving blind added multi-pesticide confirmation within 2-3 minutes. The application covers organo-phosphorous, organo-chlorine, and pyrethroid pesticides in samples such as cucumbers, tomatoes, and green peppers.

What's inside

  • Introduction
  • Highlights
  • Experimental
  • Instrument conditions
  • Sample preparation
  • Results and discussion

Frequently asked questions

What is the minimum amount of sample material required for a TSP injection?

The procedure requires placing a small amount of sample material or liquid extract into a 40 pL disposable micro-vial.

What is the maximum concentration of pesticides used in the repeatability test?

The test evaluated organo-chloro pesticide solutions at a concentration of 1.0 pg/mL.

How many continuous injections were performed to test the system's baseline stability?

The system was tested with 50 continuous injections of tomato extractions without a clean-up step.

Download PDF (1.5 MB)

More Agilent Manuals


View all Agilent manuals →