Agilent J&W FactorFour VF-WAXms GC Columns Application Note for Beer Analysis

This document describes the application of Agilent J&W FactorFour VF-WAXms GC Columns in beer analysis.


Agilent J&W FactorFour VF-WAXms GC Columns Application Note for Beer Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
1
File format
PDF
File size
164 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
8CA1C6D0E18B0B5DEC5CB1355CF42BD2

About this manual

The document provides a specific analytical method for detecting volatile compounds in beer. The technique utilizes Gas Chromatography with Flame Ionization Detection (GC-FID) to identify various alcohols, esters, and other volatiles present in the sample.

The procedure specifies the use of an Agilent J&W FactorFour VF-WAXms column (part number CP9211). The method includes detailed operating conditions for the injector, detector, carrier gas flow, and a temperature program ranging from 55 °C to 220 °C.

What's inside

  • Conditions
  • Technique
  • Column Specifications
  • Injector Settings
  • Detector Settings
  • Temperature Program
  • Sample Preparation
  • Peak Identification

Specifications

Carrier GasHelium, 0.7 bar
Injector Temperature250 °C
Split Ratio Flow30 mL/min
Detector Temperature275 °C
Column Part NumberCP9211

Frequently asked questions

What compounds are identified in the beer analysis?

The method identifies acetaldehyde, dimethyl sulfide, acetone, ethyl formate, ethyl acetate, ethanol, isobutanol, isoamyl acetate, and 3-methyl-1-butanol.

What is the temperature program for the analysis?

The oven starts at 55 °C held for 2 minutes, then ramps to 220 °C.