LC-MS/MS Determination of Alternaria Toxins in Vegetables and Fruit Beverages Application Note

This application note describes LC-MS/MS methods for the determination of Alternaria toxins in vegetables and fruit beverages.


LC-MS/MS Determination of Alternaria Toxins in Vegetables and Fruit Beverages Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
112 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
A883A97E683EA6D1E5736F1CF0F65F10

About this manual

This document details an analytical method for the qualitative and quantitative determination of five important alternaria toxins (AOH, AME, ALT, TEN, TEA) in vegetable and fruit beverage matrices. The method utilizes electrospray ionization LC-tandem mass spectrometry with positive/negative ion detection in MRM mode to achieve high sensitivity at sub-ng/mL levels.

The procedure involves sample extraction using acetonitrile-methanol-water mixtures, followed by cleanup and concentration via solid phase extraction (SPE) using Agilent's Bond Elut Plexa hydroxylated polymer. Separation is performed on an Agilent MonoChrom MS silica gel column with a pentafluorophenyl modification.

What's inside

  • Introduction
  • Method Overview
  • Sample Preparation (Solid)
  • Sample Preparation (Liquid)
  • MS Conditions
  • LC Conditions
  • Results and Discussion
  • Conclusion

Specifications

Column TypeAgilent MonoChrom MS silica gel column, 5 µm, 150 x 3 mm
Column Part NumberA2080150X030
SPE Cartridge TypeBond Elut Plexa 200 mg, 6 mL cartridge
SPE Cartridge Part Number12109206
Limit of Detection (LOD)>5 µg/L
Limit of Determination (LOQ)>10 pg/L

Frequently asked questions

What toxins does this method determine?

The method determines alternariol (AOH), alternariol monomethyl ether (AME), altenuene (ALT), tentoxin (TEN), and tenuazonic acid (TEA).

What sample matrices are covered?

The method covers liquid samples such as grape juice and vegetable beverages, as well as solid samples like fruit purées.

What SPE cartridge is recommended?

The application recommends using a Bond Elut Plexa 200 mg, 6 mL cartridge.

Download PDF (112 KB)

More Agilent Manuals


View all Agilent manuals →