Understanding Orthogonality in Reversed-Phase Liquid Chromatography for Easier Column Selection Method Development Appli

This document introduces orthogonality in reversed-phase liquid chromatography and achieves high resolution by comparing the selectivities of different Agilent Polaris and Agilent Pursuit columns.


Understanding Orthogonality in Reversed-Phase Liquid Chromatography for Easier Column Selection Method Development Appli - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
1.6 MB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
24AB3C9782D18452E9BF3384E90EC2B3

About this manual

Orthogonality in chromatography refers to alternative selectivity between separations, which is essential for addressing insufficient resolution that can mask impurities or degradation peaks. This document details a study comparing different Agilent Polaris and Pursuit stationary phases to identify pairs of columns with orthogonal retention patterns.

The research screened ten premier Polaris and Pursuit columns against a broad range of analytes under identical conditions using an Agilent ProStar HPLC System. By analyzing the scatter in retention factors, the authors identified specific column combinations that exhibit high orthogonality, helping end-users save time and effort during method development.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • HPLC Conditions and Detection
  • Results and Discussion
  • Orthogonality Measurement

Frequently asked questions

What is orthogonality in chromatography?

Orthogonality refers to alternative selectivity between separations, achieved by modifying method parameters or choosing stationary phases with different selectivities.

What is the purpose of orthogonal separations in method development?

Orthogonal or 2D separations are needed to address insufficient resolution that can mask an impurity or a sample degradation peak.

What flow rate was used for analyzing compounds at 50% organic conditions?

The analysis was performed at a flow rate of 1.0 mL/min.

Which wavelength was used for UV detection on most analytes?

UV Detection used for most analytes was 254 nm.